Interferometer for Measuring Displacement to Within 20 pm
TBMG-29379
7/1/2003
- Content
An optical heterodyne interferometer that can be used to measure linear displacements with an error -20 pm has been developed. The remarkable accuracy of this interferometer is achieved through a design that includes (1) a wavefront split that reduces (relative to amplitude splits used in other interferometers) self interference and (2) a common-optical-path configuration that affords common-mode cancellation of the interference effects of thermal-expansion changes in optical-path lengths.
- Citation
- "Interferometer for Measuring Displacement to Within 20 pm," Mobility Engineering, July 1, 2003.