Computer Generated Hologram System for Wavefront Measurement System Calibration
TBMG-9128
02/01/2011
- Content
Computer Generated Holograms (CGHs) have been used for some time to calibrate interferometers that require nulling optics. A typical scenario is the testing of aspheric surfaces with an interferometer placed near the paraxial center of curvature. Existing CGH technology suffers from a reduced capacity to calibrate middle and high spatial frequencies. The root cause of this shortcoming is as follows: the CGH is not placed at an image conjugate of the asphere due to limitations imposed by the geometry of the test and the allowable size of the CGH.
- Citation
- "Computer Generated Hologram System for Wavefront Measurement System Calibration," Mobility Engineering, February 1, 2011.