Alternative to Hydrogen/Helium as Flame Ionization Detector Fuel
Technical Paper
2013-01-1045
ISSN: 0148-7191, e-ISSN: 2688-3627
Annotation ability available
Sector:
Language:
English
Abstract
Flame ionization detector (FID) analyzers used in emission
testing to measure total hydrocarbon emissions have been operating
for the last forty years on a fuel mixture of 40% H₂ and 60%
helium. These mixtures were selected based on research studies
reported in the literature indicating that this particular mixed
fuel combination gave the best sensitivity and relative response of
the different hydrocarbons present in vehicle exhaust with respect
to propane, the calibration gas.
During the past few years, it was announced that there is a
worldwide shortage of helium which triggered the automotive
industry to look for alternatives for helium to be used in FID
fuels. Helium which is produced as a byproduct from natural gas
fields is non-renewable, expensive, and extremely rare on the
earth. Current supply cannot keep up with demand. There are only
few natural gas fields producing helium and unless new natural gas
fields are found, current helium amounts will continue to dwindle.
Estimates indicate that the private reserves may run out within the
next ten years or so. Meetings with gas suppliers over the past few
years led to the conclusion that a suitable alternative to helium
for FID Fuel must be investigated.
In this study, several alternative candidates to H₂/He were
considered: Hydrogen/Nitrogen, Hydrogen/Argon and 100% Hydrogen.
This paper discusses these different options and describes the
testing performed to evaluate these different candidates to
determine the best performing alternative fuel. Linearity, relative
response with respect to propane and 10 to 90% response time were
used to evaluate the different candidate alternative fuels. The
evaluation was performed on bag, modal and heated Emerson/Rosemount
Flame Ionization Detector (FID) analyzers. Bag dilute and modal
vehicle hydrocarbon emission data with different fuels were
collected simultaneously on modified FID analyzers operating on
H₂/N₂ and a conventional H₂/He FID analyzers for direct comparison.
A heated Rosemount hot FID was also converted to operate on H₂/N₂
and evaluated with continuous dilute HC measurement on a dilution
tunnel with a diesel correlation vehicle. The impact on mass
emissions at the standards were evaluated and discussed.
Recommended Content
Technical Paper | Dual-Fueling Diesel/NGV Technology |
Technical Paper | The Development of the GM 2.2LCNG BI-Fuel Passenger Cars |
Technical Paper | The GMR-GMR&D Fuels and Lubricants Department - Its History and Accomplishments |
Authors
Topic
Citation
Yassine, M., La Pan, M., and Nayfeh, K., "Alternative to Hydrogen/Helium as Flame Ionization Detector Fuel," SAE Technical Paper 2013-01-1045, 2013, https://doi.org/10.4271/2013-01-1045.Also In
References
- Reschke , G. Optimization of a Flame Ionization Detector for Determination of Hydrocarbon in Diluted Automotive Exhausts SAE Technical Paper 770141 1977 10.4271/770141
- Springer , K. and Dietzmann , H. Diesel Exhaust Hydrocarbon Measurement - A Flame-Ionization Method SAE Technical Paper 700106 1970 10.4271/700106
- Teague , D. , Lesniak , E. , and Loeser , E. Recommended Flame Ionization Detector Procedure for Automotive Exhaust Hydrocarbons SAE Technical Paper 700468 1970 10.4271/700468
- Sun , E. , McMahon , W. , Peterson , D. , Wong , J. et al. Oxygen Quench Effect on Flame Ionization Detector for Hydrocarbon Emission Measurements SAE Technical Paper 2004-01-1431 2004 10.4271/2004-01-1431
- Sherman , M. , Henney , T. , Akard , M. , and Lipke , W. Analysis of Flame Ionization Detector Oxygen Effects Using Blended Cylinders SAE Technical Paper 2005-01-0688 2005 10.4271/2005-01-0688
- Sherman , M. , Akard , M. , and Nakamura , H. Flame Ionization Detector Oxygen Quench Effects on Hydrocarbon Emission Results SAE Technical Paper 2004-01-1960 2004 10.4271/2004-01-1960
- http://www.independent.co.uk/news/science/why-the-world-is-running-out-of-helium-2059357.html
- 70/220/EEC Annex III Type I Test Appendix 6 (2.3) 78 Response Factors of Different Hydrocarbons and Recommended Limits April 6 1970
- 40 CFR 1065.362 Non-stoichiometric Raw Exhaust FID O 2 Interference Verification
- Bramston-Cook , Randall Primer on Flame Ionization Detectors Lotus Flower, Inc., Lotus Consulting June 4 2011